Metrohm 788 IC Filtration Sample Processor User Manual

Page 111

Advertising
background image

4.6 Standard methods

788 IC Filtration Sample Processor

103

Settings in the «IC Metrodata» program

1. Definition of anion method for channel 1 (analysis time, start pa-

rameters for IC instruments 732, 733 and 709, etc.), additional asso-

ciation of this method with the following time program for the 732/1

IC Detector:

Method / Equipment / Setup / Program / Text:

0.0 Puls_788_set Set remote line 3 at 732/1 to 1 (s. below)
0.1 Puls_788_reset Set remote line 3 at 732/1 to 0 (s. below)
0.2 Flag

end

Program end

Method / Equipment / Setup / Program / Remote configuration:

Puls_788_set =

**1***** Set remote line 3 at 732/1 to 1

Puls_788_reset =

**0*****

Set remote line 3 at 732/1 to 0

Method / Equipment:

Start with method

2. Definition of cation method for channel 2 (analysis time, start pa-

rameters for IC instruments 732, 733 and 709, etc.). The analysis

time

Duration

must be shorter than the analysis time for the anion

method.

Procedure at program start

1. Create queue with cation method for channel 2 in the program «IC

Metrodata» and start it.

2. Start 788 IC Filtration Sample Processor with <START>.
3. Create queue with anion method for channel 1 in the program «IC

Metrodata» and start it.