Metrohm 788 IC Filtration Sample Processor User Manual

Page 29

Advertising
background image

2.3 Attaching the accessories

788 IC Filtration Sample Processor

21

Do not connect the dialysis cell with anything other than the

6.2744.000 PVDF compression fittings supplied. If 6.2744.010

PEEK compression fittings are used, stress cracks may appear in the

ultra-filtration cell !

3 Connection pump tubing 10 – ultra-filtration cell

x Mount a PEEK compression fitting 6

666 on the end of the PTFE

capillary 24

24

24

24 and screw it on to coupling 19

19

19

19.

x Mount a PVDF compression fitting 22

22

22

22 on the other end of the

PTFE capillary 24

24

24

24.

x Screw this PEEK capillary with the PVDF compression fitting

to inlet opening 48

48

48

48 of the bottom part of the cell (see Fig. 9).

x Push coupling 19

19

19

19 on to the outlet end of the pump tubing 10

10

10

10

(yellow/yellow) (see Fig. 10).

4 Connection ultra-filtration cell – waste

x Mount a PVDF compression fitting 22

22

22

22 on the end of the PTFE

capillary 25

25

25

25.

x Screw this end of the PTFE capillary with the PVDF

compression fitting to outlet opening 49

49

49

49 of the bottom part of

the cell (see Fig. 9).

x Mount the other end of the PTFE capillary to a waste bottle

(e.g. optional: 6.1608.070 bottle with 6.1602.150 siphon

GL45).

5 Connection ultra-filtration cell – pump tubing 11

x Mount a PEEK compression fitting 6

666 on the end of the PTFE

capillary 21

21

21

21 and screw it on to coupling 19

19

19

19.

x Mount a PVDF compression fitting 22

22

22

22 on the other end of the

PTFE capillary 21

21

21

21.

x Screw this PTFE capillary with the PVDF compression fitting

to outlet opening for the filtrate 47

47

47

47 of the top part of the cell

(see Fig. 9).

x Push coupling 19

19

19

19 on to the inlet end of the pump tubing 11

11

11

11

(orange/yellow) (see Fig. 10).

6 Connection pump tubing 11 – injection valve

(Underlined numbers are parts or controls from the Instructions

for Use for the 733 IC Separation Center)
x Mount a PEEK compression fitting 6

666 on the end of the PEEK

capillary 20

20

20

20 and screw it on to coupling 19

19

19

19.

x Push coupling 19

19

19

19 on to the outlet end of the pump tubing 11

11

11

11

(orange/yellow) (see Fig. 10).

x At the 733 IC Separation Center, loosen the rotary nipple

screwed onto the interior side of connection 22

22

22

22 or 28

28

28

28.

x Take PTFE suction tubing 84

84

84

84 (see Fig. 14 and Fig. 15 of the

732/733 Instructions for Use) completely out of connection 22

22

22

22

or 28

28

28

28 and unscrew from connection "1" of injection valve 66

66

66

66.