2 Installation

788 IC Filtration Sample Processor

22

x Pull the PEEK capillary 20

20

20

20 through the opening 22

22

22

22 or 28

28

28

28 of

the 733 IC Separation Center and screw onto connection "1"

of injection valve 66

66

66

66 using a 6.2744.010 PEEK compression

fitting 6

666.

x Retighten rotary nipple on the interior side of connection 22

22

22

22

or 28

28

28

28 to fix the capillary 20

20

20

20.

7 Tubing connection injection valve – waste

x Insert 6.2744.020 coupling (from 733 accessories) into

connection 21

21

21

21 or 27

27

27

27 of the 733 IC Separation Center.

x Screw PTFE suction tubing 84

84

84

84 onto the 6.2744.020 coupling

attached to connection 21

21

21

21 or 27

27

27

27 and lead it into the waste

container.

In the case of the 733.0020 IC Separation Center with two injection

valves, it is possible to fill both sample loops from the same 788 IC

Filtration Sample Processor. For this, connection "1" of valve A (outlet

of the sample loop) must be connected to connection "2" of valve B

(inlet of the sample loop) using a 6.1803.040 PEEK capillary (15 cm).

2.3.10 Tubing connections 788 – 754 Dialysis Unit

If the 788 IC Filtration Sample Processor is used for an IC system with

sample dialysis (see section 2.4.7 or section 2.5.8), the peristaltic pump

at the 754 Dialysis Unit is used for sample conveying instead of the

pump at the 788 IC Filtration Sample Processor. Additionally, the ultra-

filtration cell is bridged. The tubing connections between the 754

Dialysis Unit and the 733 IC Separation Center have to be made as

shown in Fig. 8 of the 754 Instruction for Use. The only change concerns

the inlet of the sample, which is installed as follows:

Install a PEEK compression fitting 6.2744.010 on the inlet end of the

PTFE tubing 26

26

26

26 (6.1803.030, see Fig. 8 of the 754 Instruction for Use)

connected to the 754 Dialysis Unit and screw it on the PEEK

compression fitting 5

555 (see Fig. 6).

2.3.11 Fixing tubing and cables

In order to fix tubing or cables in the

guide chain 9

999 any chain link may be

opened with a screw driver or another

appropriate tool..

Fig. 12: Opening chain links