6.1 Error messages

788 IC Filtration Sample Processor

133

6 Appendix

6.1

Error messages

When an error occurs the execution of the active command is inter-

rupted and an error message is displayed (display blinks). This must be

acknowledged with the <QUIT> key.

If an error occurs during processing of a sample series, the 788 IC Fil-

tration Sample Processor will then be switched into the "HOLD" state.

After the cause of the error has been rectified, the sample series can be

continued with the next command in the active sequence by pressing

the <START> key. If the error cannot be eliminated, the method run-

ning can also be halted with <STOP>.

The list of possible error messages and their causes:

Error message

Cause

* battery low

The battery for the permanent storage

of user methods must be replaced.

* wrong rack

The rack positioned is not the one

that was assigned to the method un-

der

'parameters'

.

* rack data missing

No sample rack is in position or no

rack data can be found for the sam-

ple rack that is in place.

* raise lift first

Turning of a rack could not be carried

out because the lift was below the

defined shift position.

* user memory full

The memory for the user-defined

methods is full. Before saving a new

method, methods that are not used

or used only rarely must be deleted.