2.3 Attaching the accessories

788 IC Filtration Sample Processor

17

Use only ultrapure water or ethanol to clean the dialysis cell;

organic solvents (e.g. acetone) will cause damage to the plexiglas

cell!


2 Prepare filtration membrane

x Use the 6.2831.010 tweezers to extract a new filtration

membrane 50

50

50

50 (6.2714.020) from its packaging and immerse

in a Petri dish containing ultra pure water for approx. 2 min

until the membrane is completely saturated with water.

Because the filtration membrane it is stretched out thereby, it has to

be watered before inserting into the ultra-filtration cell. Otherwise, the

membrane could be crumbled in the cell, which leads to blocking.

3 Insert filtration membrane

x Lay top part 43

43

43

43 on a paper wipe with its inside facing

upwards.

x Insert sealing ring 44

44

44

44 in the appropriate recess in the top part.

x Using the tweezers, place the wet dialysis membrane inside

the sealing ring on the top part of the filtration cell.

Do not connect the dialysis cell with anything other than the

6.2744.000 PVDF compression fittings supplied. If 6.2744.010

PEEK compression fittings are used, stress cracks may appear in the

ultra-filtration cell !

4 Close ultra-filtration cell

x Insert sealing ring in the appropriate recess in top part 43

43

43

43.

x Place top part on bottom part 45

45

45

45 so that both parts lie flush.

x Using the 6.2621.070 Allen key, screw the 5 screws 46

46

46

46 right

in and tighten up.

5 Attach ultra-filtration cell

x Insert the assembled ultra-filtration cell in the cell holder 26

26

26

26

according to Fig. 1 or Fig. 11. The heads of the five screws

46

46

46

46 fit to the small holes of the cell holder.