Metrohm 788 IC Filtration Sample Processor User Manual

Page 37

Advertising
background image

2.4 Installation for full control with «IC Net»

788 IC Filtration Sample Processor

29

Settings in the «IC Net 2.1» program
1. Definition of new program parameters for the 788 IC Filtration

Sample Processor under Configuration/Control:

FILL A 2

= ---***0*010***

INJECT A 2

= ---***1*000***

ZERO 2

= ---***0*011***

2. Program for the 788 IC Filtration Sample Processor:

001 Ctrl INIT
002 Move

sample

003 Lift work
004 Ctrl ZERO

1

005

Ctrl

FILL A 1

006

Ctrl

STEP MSM 753

007

Pump 240 s

008

Ctrl

INJECT A 1

009 Move

sample+1

010 Lift work
011 Ctrl ZERO

2

012 Ctrl Init

732

013 Scan

Wait1

014

Ctrl

FILL A 2

015 Ctrl Init

732

016 Pump

240

s

017

Ctrl

INJECT A 2

018 Ctrl Init

732

 Initialize remote interface
 Move needle to sample position
 Place lift with needle to working position
 Trigger autozero at 732/1 IC Detector
 Switch injection valve A at 733 to "Fill"
 Switch 753 suppressor module to next position
 Fill sample loop A with sample during 240 s
 Switch injection valve A at 733 to "Inject"
 Move needle to sample position (+1)
 Place lift with needle to working position
 Trigger autozero at 732/2 IC Detector
 Initialize remote lines at 732/1 and 732/2
 Wait until 732/1 IC Det. sends signal on remote line 3
 Switch injection valve B at 733 to "Fill"
 Initialize remote lines at 732/1 and 732/2
 Fill sample loop B with sample during 240 s
 Switch injection valve B at 733 to "Inject"
 Initialize remote lines at 732/1 and 732/2

3. Definition of new program parameters for the 723/1 IC Detector

under Programm/Remote Configuration:

Start_788

=

**1*****

Reset_Start_788

=

**0*****

4. To synchronize the second injection to the start of the second

chromatogram, a time program is started for the 732/1 IC Detector

directly with the first injection:

1.0 Start_788

1.1

Reset_Start_788

1.2

Flag end

Both detectors 732/1 and 732/2 are registered in the system window

under Setup/Start mode in the field 'Start with inject' .

4. By this procedure, with a delay of one minute, the 788 IC Filtration

Sample Processor receives a signal for processing the second

sample. The total resulting delay (5 min.) is now independent from

the time the sampler needs for rotating to the next sample.

Therefore, the method for the second sample injection (e.g. method

'cation') gets a start delay under Method Setup/Measure:

Start delay:

5.0 min