2.3 Attaching the accessories

788 IC Filtration Sample Processor

13

2.3.4

Installing the splash protection

To avoid any danger of injury by the needle, the 6.2751.040 splash

protection must always be installed when operating the 788 IC

Filtration Sample Processor!

1 Remove holding screws

Remove the holding screws 37

37

37

37 and the washer mounted on the

screw threads 12

12

12

12 at tower 27

27

27

27 using the 6.2621.100 Allen key.

2 Remove protective film from splash protection

Remove the plastic film glued on both sides of the splash

protection 1

111.

3 Install splash protection

Attach splash protection onto the screw threads at the tower and

fix it with the holding screws and the washer using the

6.2621.100 Allen key.

37

37

37

37

37

37

37

37

1111

Fig. 5: Installing the splash protection

2.3.5

Installing the needle or tubing

1 Remove PEEK compression fitting 5

Remove PEEK compression fitting 5

555 screwed onto the needle

holder 4

444.

2 Insert needle

Insert needle 2

222 (6.1835.010) PEEK needle or 6.1835.020 PEEK

tubing) completely into the opening of the needle holder 4

444.

3 Fix needle or tubing

Tighten compression fitting 5

555 in needle holder 4444 by hand

(never use tools!).