4 Operation

788 IC Filtration Sample Processor

104

4.6.5 Method "Preconc"

Application

Preconcentration of samples with very low concentrations in an precon-

centration column (trace analysis). This method is not planned for us-

age together with the ultra-filtration cell. However, if ultra-filtration is

done, the preconcentration time in line 8 has to be increased to main-

tain a sufficient filtering time, but without complete consuming of the

sample solution.

Interconnection

Siehe section. 2.6.3.

Installation of the preconcentration column

The preconcentration column, either consisting of the 6.1006.200 Met-

rosep Anion preconcentration cartridge and the 6.2828.010 glass car-

tridge holder or the 6.1006.300 preconcentration column, is installed at

the injection valve of the 733 IC Separation Center instead of the sam-

ple loop (see Fig. 25). It is important to install the sample inlet from the

788 IC Filtration Sample Processor at connection 2 of the injection valve

in order to run the preconcentration column for the enrichment and the

sample injection in the opposite direction.

1

2

4

5

3

6

1

2

4

5

3

6

Fig. 25: Installation of the preconcentration column

Waste

Sample

Sample

Waste

Sample injection

Preconcentration

IC column

Preconc. column

Eluent

IC column

Eluent