Metrohm 788 IC Filtration Sample Processor User Manual

Page 113

Advertising
background image

4.6 Standard methods

788 IC Filtration Sample Processor

105

Program of 788 IC Filtration Sample Processor

788 IC Filtr. Proc. 5.788.0010
parameters
method Preconc
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 CTL:Rm: INJECT A 1
3 CTL:Rm: PUMP 752 ON
>sample sequence
1 SCN:Rm : Wait1
2 SCN:Rm : Pump1 ?
3 MOVE 1 : sample
4 LIFT: 1 : work mm
5 PUMP 1.1 : ON s
6 WAIT 60 s
7 CTL:Rm: FILL A 1
8 WAIT 120 s

9 PUMP 1.1 : OFF s
10 CTL:Rm: ZERO 1
11 CTL:Rm: INJECT A 1
>final sequence
1 WAIT 1200 s
2 CTL:Rm: PUMP 752 OFF
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------

 Report header with serial number and program ver-

sion

 Method name
 Number of samples to be processed (entire sample

rack)

 Initialize remote interface
 Switch injection valve A at 733 to "Inject"
 Switch 752 Pump Unit on for suppression


 Waiting until 732 IC Det. sends signal on rem. line 3
 Waiting until 709 IC Pump runs
 Move needle to sample position
 Place lift with needle to working position
 Switch on pump at 788
 Waiting time 60 s
 Switch injection valve A at 733 to "Fill"
 Sample preconcentration during 120 s (preconcen-

tration time must be adapted)

 Switch off pump at 788
 Trigger autozero at 732 IC Detector
 Switch injection valve A at 733 to "Inject"


 Wait for final regeneration of suppressor
 Switch 752 Pump Unit off
 —— Settings for changer functions ——




 —— Reaction to manual stop ———

Settings in the «IC Metrodata» program

Definition of IC method (analysis time, start parameters for IC instru-

ments 732, 733 and 709, etc.), additional association of this method

with the following time program for the 732 IC Detector:

Method / Equipment / Setup / Program / Text:

0.0 Puls_788_set Set remote line 3 at 732/1 to 1 (s. below)
0.1 Puls_788_reset Set remote line 3 at 732/1 to 0 (s. below)
0.2 Flag

end

Program end

Method / Equipment / Setup / Program / Remote configuration:

Puls_788_set =

**1***** Set remote line 3 at 732/1 to 1

Puls_788_reset =

**0*****

Set remote line 3 at 732/1 to 0

Method / Equipment:

Start with method

Procedure at program start

1. Start 788 IC Filtration Sample Processor with <START>.
2. Create new queue in the program «IC Metrodata» and start it.