Page 20: 3 installing the plug cover

Advertising
background image

2 Installation

788 IC Filtration Sample Processor

12

2.3

Attaching the accessories

For attaching the accessories at the 788 IC Filtration Sample

Processor, proceed in the order described below.

2.3.1

Connecting the swing head

Plug in the branch plug of the connection cable 29

29

29

29 permanently

attached to the swing head into the remote connection socket at the

788 IC Filtration Sample Processor and screw it onto this connection

using a screw-driver (see Fig. 1).

2.3.2

Connecting the keyboard

1

2

Connecting the keyboard

Connect the 6.2142.010 keyboard

to the keyboard connection 39

39

39

39

"Keyboard". For disconnection,

press the plug together slightly on

both sides.

Switch on instrument

Switch on the 788 IC Filtration

Sample Processor with mains

switch 34

34

34

34. The keyboard display

lights up. The instrument is

initialized and the lift is raised

completely.

Fig. 4: Keyboard connection

2.3.3

Installing the plug cover

To prevent any contamination of the mains and remote connection by

spilled solvents or chemicals, the 6.2752.010 plug cover must always

be installed when operating the 788 IC Filtration Sample Processor!

Install the plug cover 31

31

31

31 in the corresponding guide groove above

mains connection plug 33

33

33

33 and remote connection 30

30

30

30 (see Fig. 1 and

Fig. 2).