2 Installation

788 IC Filtration Sample Processor

10

2.2 Mains

connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with the mains voltage set wrongly and/or

wrong mains fuse there is a danger of fire!


2.2.1

Setting the mains voltage


34

34

34

34

33

33

33

33
32

32

32

32

220
---- 240
V

100

---- 120

V

220

---- 240

V

100
---- 120
V

Fig. 3: Setting the mains voltage

Before switching on the 788 IC Filtration Sample Processor for the first

time, check that the mains voltage set on the instrument (see Fig. 3)

matches the local mains voltage. If not, reset the mains voltage on the

instrument as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 33

33

33

33 of the

788 IC Filtration Sample Processor.

2 Remove fuse holder

Using a screwdriver, loosen fuse holder 32

32

32

32 beside the mains

connection and take out completely.

3 Check fuse

Carefully take the fuse installed for the desired mains voltage out

of fuse holder and check its specifications (the position of the

fuse in the fuse holder is marked by the white arrow imprinted

next to the mains voltage range):

100}

}

}

}120 V 0.5 A (slow-blow) Metrohm-Nr.

U.600.0013

220}

}

}

}240 V 0.25 A (slow-blow) Metrohm-Nr.

U.600.0010

220 – 240 V

100 – 120 V

32

32

32

32 Fuse holder

33

33

33

33 Mains connection

plug

34

34

34

34 Mains switch