6.2 Technical data

788 IC Filtration Sample Processor

135

6.2 Technical

data

Lift

Lift path

ca. 125 mm

Load

ca. 30 N

Stroke speed

adjustable, 3}12 mm/s

Turntable

Turning speed adjustable,

3}20 angular degrees/s

Pump

Pump type

2-channel peristaltic pump with rotational speed

of 20 /min

Pump capacity

(with water, without

counterpressure; depends

on contact pressure )

with 6.1826.010 pump tubing:

typ. 1.5}1.7 mL/min
with 6.1826.030 pump tubing:

typ. 0.5}0.6 mL/min
with 6.1826.070 pump tubing:

typ. 1.8}2.0 mL/min

Pressure

max. 1.5 bar (0.15 MPa)

Pump tubing material

PVC (Tygon

£

)

Ultra-filtration cell

Material

Plexiglass (polymethyl methacrylate)

Solvent compatibility

Water, ethanol

(no organic sovents!)

Cell volume

2 x 210 PL

Diameter of membrane

47 mm

Parts and controls

LCD-Display

2 lines of 24 characters each, height 5 mm

Keyboard

chemically resistant membrane keypad made of

polyester with function and numeric keys

Mains connection

Voltage

100...120 V

220...240 V

Switchable with voltage selector in fuse holder

(see section 2.2.1)

Frequency

50...60 Hz

Power consumption

40 VA

Fuse

5 mm Ø, 20 mm length

100}120 V: 0.5 A (slow-blow)

220}240 V: 0.25 A (slow-blow)

Advertising