Metrohm 788 IC Filtration Sample Processor User Manual
Page 34
2 Installation
788 IC Filtration Sample Processor
26
2.4.4 MIC
1
The 788 IC Filtration Sample Processor is connected to an IC system
with electronic suppression consisting of 732 IC Detector, 733 IC
Separation Center, 709 IC Pump and 762 IC Interface as shown in
Fig. 13 using the 6.2141.110 remote connection cable and the
6.2134.080 RS232 cable.
732
733
709
762
PC
6.2134.100
6.2128.130
6.2115.070
6.2134.090
6.2125.090
788
6.2134.080
6.2141.110
Fig. 13: Interconnection with anion or cation IC system
with electronic suppression (MIC 1)
Settings in the «IC Net 2.1» program
Creation of a time program for the 788 IC Filtration Sample Processor:
001 Ctrl
INIT
732
002 Move sample
003 Lift
work
004 Ctrl
ZERO
1
005 Ctrl
FILL A 1
006 Pump 240 s
007 Ctrl
INJECT A 1
Initialize remote interface at 732
Move needle to sample
Place lift with needle to working position
Trigger autozero at 732
Switch injection valve A at 733 to "Fill"
Fill sample loop with sample during 240 s
Switch injection valve A at 733 to "Inject"
Modular anion or cation system
with electronic suppression