Metrohm 788 IC Filtration Sample Processor User Manual

Page 34

Advertising
background image

2 Installation

788 IC Filtration Sample Processor

26

2.4.4 MIC

1


The 788 IC Filtration Sample Processor is connected to an IC system

with electronic suppression consisting of 732 IC Detector, 733 IC

Separation Center, 709 IC Pump and 762 IC Interface as shown in

Fig. 13 using the 6.2141.110 remote connection cable and the

6.2134.080 RS232 cable.

732

733

709

762

PC

6.2134.100

6.2128.130

6.2115.070

6.2134.090

6.2125.090

788

6.2134.080

6.2141.110

Fig. 13: Interconnection with anion or cation IC system

with electronic suppression (MIC 1)

Settings in the «IC Net 2.1» program


Creation of a time program for the 788 IC Filtration Sample Processor:

001 Ctrl

INIT

732

002 Move sample
003 Lift

work

004 Ctrl

ZERO

1

005 Ctrl

FILL A 1

006 Pump 240 s
007 Ctrl

INJECT A 1

 Initialize remote interface at 732
 Move needle to sample
 Place lift with needle to working position
 Trigger autozero at 732
 Switch injection valve A at 733 to "Fill"
 Fill sample loop with sample during 240 s
 Switch injection valve A at 733 to "Inject"

Modular anion or cation system
with electronic suppression