4.6 Standard methods

788 IC Filtration Sample Processor

107

Settings in the «IC Metrodata» program

Definition of IC method (analysis time, start parameters for IC instru-

ments 732, 733 and 709, etc.), additional association of this method

with the following time program for the 732 IC Detector:

Method / Equipment / Setup / Program / Text:

10.0 Puls_788_set Set remote line 3 at 732/1 to 1 (s. below)
10.1 Puls_788_reset Set remote line 3 at 732/1 to 0 (s. below)
10.2 Flag end

Program end

Method / Equipment / Setup / Program / Remote configuration:

Puls_788_set =

**1***** Set remote line 3 at 732/1 to 1

Puls_788_reset =

**0***** Set remote line 3 at 732/1 to 0

Method / Equipment:

Start with inject

The time of 10 min. given above is the starting time for the dialysis of

the next sample. It results from the total chromatogram time (e.g. 20

min.) minus 10 min. Therefore, these time values have to be adapted to

varying chromatogram times.

Procedure at program start

1. Create new queue in the program «IC Metrodata» and start it.

2. Start 788 IC Filtration Sample Processor with <START>.