Metrohm 788 IC Filtration Sample Processor User Manual

Page 43

Advertising
background image

2.5 Installation with compact IC systems

788 IC Filtration Sample Processor

35


The corresponding settings in the PC software vary according to the IC

system which is used:

Settings in the «761 Compact IC» program

A time program must be drawn up for the selected system which first

produces an impulse at remote lead 3 to start sample changing on

the 788 IC Filtration Sample Processor and then switches the injection

valve to the "Fill" position. A further impulse is then produced to start

filling the sample loop within 240 s. At the end the injection valve is

switched to the "Inject" position, which also starts data acquisition.

Run conditions

In order that the joint operation of 761 – 788 functions properly the

following conditions must be fulfilled:
x The program must be started first on the 788 IC Filtration Sample

Processor, then the "Sample Queue" on 761 Compact IC.

x Remote lead 3 on the Compact IC must be set to 0 at the start of

each determination (Set

System startup values

:

Remote line 3 = 0

).