2.3 Attaching the accessories

788 IC Filtration Sample Processor

15

35

35

35

35

42

42

42

42

41

41

41

41

2222

Fig. 8: Adjusting the sample rack

2 Move to adjusting position

Press <MOVE>, enter number '37' and confirm with <ENTER>.

Sample rack and swing head are turned until needle is above

the adjusting position 37 (first opening of the middle row).

3 Check needle position

x Press <

Ð

> until the needle is ca. 1 cm above the upper level

of the sample rack.

x Check needle position: If the needle cannot be lowered

unhindered through the upper hole of the adjusting position,

continue directly with point 4.

x Continue lowering the needle by pressing <

Ð

> until the

needle is ca. 1 cm above the lower level of the sample rack.

x Check needle position: If the needle cannot be lowered

unhindered through the lower hole of the adjusting position,

continue directly with point 4.

x Lower needle completely by pressing <

Ð

>.

x Check needle position: If the needle is in the middle of the

lower hole, the sample rack must not be adjusted (continue in

this case with point 5).

4 Adjust sample rack

x Loosen the three adjusting screws 42

42

42

42 on the lower level of the

sample rack using the 6.2621.100 Allen key.

x Carefully turn the two upper levels of the sample rack by hand

until the lowered needle is exactly in the middle of the lower

hole at the adjusting position 41

41

41

41.

x Tighten the adjusting screws.

5 Move to rest position

Press <RESET> to move the sample rack to the rest position.

2222 PEEK needle (6.1835.010) or

PEEK tubing (6.1835.020)

35

35

35

35 Sample rack (6.2041.430)

41

41

41

41 Adjusting position 37

42

42

42

42 Adjusting screw