4 Operation

788 IC Filtration Sample Processor

98

All methods with the annex 'Seg' only differ by one additional step:

PUMP 1.1 : 5 s

This command is processed before the lift of the 788 IC Filtration Sam-

ple Processor is placed to working position. Thereby, a small segment

of air is introduced into the tubing between two samples. Therefore,

these methods should not be used together with the ultra-filtration cell.
In the following methods, this command is written in brackets without a

line number. The corresponding methods with the annex 'Seg' contain

this line regularly with its one line number, followed by increased line

numbers for the next lines.
It is recommended to work through every new method step-by-step with

the TRACE function, making adjustments where necessary, before

starting the method for the first time.
The following description of the most standard methods is based on

the usage of the PC software «IC Metrodata» for simple remote control

of the IC system. Certainly, the program «IC Net 2.X» can be used in-

stead. Additionally, the programming of the 732 IC detector is possible

by direct editing at the device.