2 method parameters – <param> key – Metrohm 788 IC Filtration Sample Processor User Manual

Page 84

Advertising
background image

4 Operation

788 IC Filtration Sample Processor

76

4.3.2

Method parameters – <PARAM> key

PARAM

The <PARAM> key combines the most important parameter settings

for the 788 IC Filtration Sample Processor. All these settings constitute

a method and may be saved as such (see section 4.3.8). The key

opens the following main menu:

parameters
number of samples: rack

Number of samples to be processed

(parameter)

parameters
>start sequence

Start sequence before sample series

parameters
>sample sequence

Sample sequence for sample series

parameters
>final sequence

Final sequence after sample series

parameters
>changer settings

Sample changer settings

parameters
>manual stop

Options with manual stop

Repeated pressing of the <PARAM> key selects the submenus in turn.

Use <Ð> or <Ï> to move up or down one menu item; use

<HOME> or <END> to move to the first or last menu item respec-

tively. The submenu is opened with <ENTER>, <QUIT> returns to the

normal state. The following listings show all dialog items which appear

under <PARAM>. The values shown in the display are the default val-

ues, the possible entry values or ranges are shown below the display.

parameters
number of samples: rack

1}999,rack,*

Number of samples to be processed

rack

= all sample positions of the rack

*

= infinite number of samples

All sample positions of an engaged rack will be proc-

essed when the instrument is on the

'rack'

setting

(max. number of rack positions – number of special

beakers defined). It is important that the 788 IC Filtration

Sample Processor can recognize the rack. This is only

possible when the rack is at the ground position. It is

recommended to initialize the 788 IC Filtration Sample

Processor with the <CLEAR> key or <ENDSEQ> and

<ENTER> after every rack change.

Advertising