6.5 Initialize data memory

788 IC Filtration Sample Processor

149

x Prepare instrument for diagnostic test (see section 6.4.2).

x If need be, press <9> key repeatedly until

diagnosis

>RAM initialization

x Press <ENTER> key to open the following diagnostic menu:

>RAM initialization
select: param

Set method parameters with default values.

>RAM initialization
select: config

Set config parameters with default values.

>RAM initialization
select: setup

Set setup parameters with default values.

>RAM initialization
select: assembly

Set assembly parameters with default values.

>RAM initialization
select: all

Set all above parameters with default values

and deletes all user-defined methods.

x Pressing the <SELECT> key selects the submenus in turn. The indi-

vidual initialization alternatives are accessed using the <ENTER>

key, exit is with the <QUIT> key.

x The table shows which parts of the data memory are affected by the

corresponding initialization alternatives. When the instrument reacts

with a system crash after being switched on (undefined display, no

reactions to keystroke, etc.) we recommend to carry out the initializa-

tion with the submenu

'all'

.

x If need be, press the <SELECT> key repeatedly until:

>RAM initialization

select: all

x Press <ENTER>.

diagnosis

>RAM test

x Press <QUIT>.

The instrument quits the diagnostic menu and runs a power on reset.