3 selecting a suitable filtration membrane – Metrohm 788 IC Filtration Sample Processor User Manual

Page 117

Advertising
background image

4.7 Filtration

788 IC Filtration Sample Processor

109

Finally, the sample matrix and the specifications of the analysis method

determine the time for changing the filtration membrane. From our ex-

perience, suspensions cause membrane clogging earlier than bigger

particles in the sample matrix. The latter should be rinsed across the

membrane surface.
An instruction for changing the filtration membrane is given in section

6.3.4.

4.7.3 Selecting a suitable filtration membrane

It is possible to transfer existing procedures for sample pretreatment to

the ultra-filtration of the 788 IC Filtration Sample Processor. For this

purpose different filtration membranes can be used. Please note that

undesirable results can be obtained although membrane pore sizes are

selected according to known particle sizes of the sample.
Own investigations have shown that filtration results of common filtra-

tion membranes do not always automatically fit to their specified pore

sizes. In the following table some qualitative results of own filtration

tests are given. Aqueous suspensions with silica particles of known

sizes have been filtered on the 788 IC Filtration Sample Processor using

membranes with different nominal pore sizes.

test sample:

silica particles in water

pore size of

filtration membrane

1

result

0.5 %, 5 µm

0.15 µm

no breaking-through

0.5 %, 5 µm

3 µm

no breaking-through

0.5 %, 5 µm

8 µm

no breaking-through

0.5 %, 5 µm

10 µm

breaking-through

2

0.5 %, 5 µm

12 µm

no breaking-through

0.5 %, 1.5 µm

0.15 µm

no breaking-through

0.5 %, 1.5 µm

3 µm

breaking-through

(1)

Nominal pore size specified by manufacturer.

(2)

Except this one, all other filtration membranes are produced by a single

manufacturer.

Due to their low thickness, filtration membranes may provide lower filtra-

tion efficiency compared to filters with the same nominal pore size but

higher thickness. This should be taken into account for selection of a

suitable membrane for online sample pretreatment using the 788 IC Fil-

tration Sample Processor.