2.4 Installation for full control with «IC Net»

788 IC Filtration Sample Processor

31

Settings in the «IC Net 2.1» program

There are two ways to operate an automated Metrohm IC system with

dialysis:
A) For each sample, every analysis procedure contains first a dialysis

step followed by the chromatographic separation.

Ÿ The total analysis time per sample results from summation of

the single times of dialysis and separation.

B) Chromatographic separation of one sample and dialysis of the next

sample is done simultaneously.

Ÿ The total analysis time per sample is reduced to the time of

chromatographic separation.


Both techniques are described on the following pages.

A) Dialysis and separation of one sample in a single procedure
1. Definition of new program parameters for the 788 IC Filtration

Sample Processor under Configuration/Control:

PUMP 754 on = ---****1******

PUMP 754 off = ---****0******

2. Program for the 788 IC Filtration Sample Processor:

001 Move

sample

002 Lift work
003

Ctrl

PUMP 754 on

004 Wait 120

s

005

Ctrl

FILL B 1 / STEP

006 Wait 480

s

007 Ctrl ZERO

1

008

Ctrl

STEP MSM 753

009

Wait 120 s

010

Ctrl

FILL A 1

011

Ctrl

INJECT B 1

012 Wait 30

s

013

Ctrl

INJECT A 1

014

Ctrl

PUMP 754 off

 Move needle to sample position
 Place lift with needle to working position
 Start 754 Pump Unit
 Rinse dialysis cell with sample solution
 Stop of acceptor solution
 Dialysis time
 Trigger autozero at 732 IC Detector
 Switch 753 suppressor module to next position
 Additional dialysis time (total 600 s)
 Switch injection valve A at 733 to "Fill"
 Transfer acceptor solution to sample loop
 Fill sample loop with acceptor solution for 30 s
 Switch injection valve A at 733 to "Inject"
 Stop 754 Pump Unit