5 Interfaces

788 IC Filtration Sample Processor

114

CTL commands

Parameter Bit

pattern

Function

Signal

INIT 00000000000000

initializes the remote interface

static

INIT 732

***0000*000**0

initializes the remote lines at 732/1 and 732/2

static

PROG R/S 1

***000*******1

starts/stops time program at 732/1

pulse (200 ms)

PROG R/S 2

******0*100***

starts/stops time program at 732/2

pulse (200 ms)

PUMP R/S 1

***001*******0

starts/stops 709/1 IC Pump

pulse (200 ms)

FILL A 1

***010*******0

switches valve A at 733/1 to "Fill"

pulse (200 ms)

INJECT A 1

***100*******0

switches valve A at 733/1 to "Inject"

pulse (200 ms)

FILL B/STEP 1 ***001*******1

switches valve B at 733/1 to "Fill"

pulse (200 ms)

INJECT B 1

***110*******0

switches valve B at 733/1 to "Inject"

pulse (200 ms)

ZERO 1

***011*******0

triggers autozero at 732/1

pulse (200 ms)

PUMP 752 ON

************1*

switches on pump at 752/753

static

PUMP 752 OFF

************0*

switches off pump at 752/753

static

STEP MSM 753

***********1**

switches 753 suppressor module to next position

pulse (200 ms)

[PUMP R/S 2 ] ******0*001***

starts/stops 709/2 IC Pump

static

[FILL A 2 ] ******0*010***

switches valve A at 733/2 to "Fill"

static

[INJECT A 2 ] ******1*000***

switches valve A at 733/2 to "Inject"

static

[FILL B/STEP 2] ******0*101***

switches valve B at 733/2 to "Fill"

static

[INJECT B 2 ] ******1*010***

switches valve B at 733/2 to "Inject"

static

[ZERO 2 ] ******0*011***

triggers autozero at 732/2

static

[PUMP 754 ON ] *******1******

switches on pump at 754 (with 6.2143.220 cable)

static

[PUMP 754 OFF ] *******0******

switches off pump at 754 (with 6.2143.220 cable)

static

SCAN commands

Parameter Bit

pattern

Function

Ready1 *******1

waits until the "ready" state of 732/1 is reached

End1 ****1***

waits until the 732/1 sends the EOD signal

End2 *1******

waits until the 732/2 sends the EOD signal

Wait1 *****1**

waits until the 732/1 sets the remote line 3 to 1

Wait2 ***1****

waits until the 732/2 sets the remote line 3 to 1

Wait* ***1*1**

waits until the 732/1 and 732/2 set their remote line 3 to 1

Pump1 ?

******1*

waits until the 709/1 IC Pump is running

Pump2 ?

**1*****

waits until the 709/2 IC Pump is running

Pump* ?

**1***1*

waits until the 709/1 and 709/2 IC Pumps are running