Backup amplifier monitoring – ATEIS IDA8C User Manual

Page 102

Advertising
background image

Ateis Studio

102

© 2012 ATEÏS

Faults

Amplifier Gain Too High

If Measure > (Unloaded Reference x (100% + Tolerance%)), an Amplifier Gain Too High fault is

recognized. once this fault is happen, do the following checks:
o Make sure there is good connection between speaker and IDA8C/S.
o Ensure the speaker is a good one.
o Check gain of amplifier, maybe the value is changed after get reference.
If all check passed but still get a amplifier too high fault, change amplifier.

Amplifier Gain Too Low

If Measure < (Full Unloaded Reference x (100% - Tolerance%)), an Amplifier Gain Too Low fault

is recognized. once this fault is happen, do the following checks:
o Check if the signal goes into amplifier is too large, some amplifier has an indicator "Overload",

you can check this indicator.

o Check if speaker is good or damaged.
If all check passed but still get a amplifier too low fault, change amplifier.

If this fault generated after speaker configuration changed, you can do a get reference again to

get new values for reference.

Amplifier Fail

Measure < (Unloaded Reference / 10), once this fault is happen, do the following checks:
o Check if amplifier is power on or not.
o Check if the gain of amplifier is set to 0, if it is, set to a suitable value.
o Check if the wires between amplifier and IDA8 are good connection.
If all check passed but still get a amplifier fail fault, change amplifier.

4.1.6.6 Backup Amplifier Monitoring

Settings

First, in Ateis Studio, right click on IDA8C/S block in [Devices] window:

Advertising
This manual is related to the following products: