5-2

IM 04L51B01-01EN

DC Voltage Measurement

CH1

Hi

Lo

DC voltage/current standard

Wiring to the first channel

of the module to calibrate

(Channel ��01)

+

Temperature Measurement Using an RTD

b

B

A

CH1

H
L
G

Resistance standard

The resistance of three

lead wires must be equal.

Wiring to the first channel

of the module to calibrate

(Channel ��01)

Temperature Measurement When Using a Thermocouple

+


CH1

Copper wires

Thermocouple wires or TC extension wires

Thermocouple

wires

Copper

wires

DC voltage standard

(0°C standard temperature device ZC-114/ZA-10 by Coper Electronics)

Uses the reference junction compensation

function of the GX/GP. See page 1-15

RJC of TC Input

As the measurement terminal of the GX/GP is generally at room temperature, the actual

output of the thermocouple is different from the values given on the thermoelectromotive

force table based on 0°C. The GX/GP performs compensation by measuring the

temperature at the input terminal and adding the corresponding thermoelectromotive force to

the actual output of the thermocouple. Therefore, when the measurement terminal is shorted

(equivalent to the case when the detector tip is 0°C), the measured value indicates the

temperature of the input terminal.

When calibrating the GX/GP, this compensation voltage (thermoelectromotive force of 0°C

reference corresponding to the input terminal temperature) must be subtracted from the

output of the standard generator before application. As shown in the figure, by using the 0°C

standard temperature device to compensate the reference junction at 0°C, you can input the

thermoelectromotive force of 0°C reference from the DC voltage standard and perform the

calibration.

5.1 Maintenance