Yokogawa Touch Screen GP20 User Manual

Page 47

Advertising
background image

1-33

IM 04L51B01-01EN

Configuring the GX/GP

1

2

3

4

5

App

Note

• Numeric display of analog input channels

If a measured value of an analog input channel is over range (see below), the measured

value is indicated as “+Over” or “–Over.” If a burnout is detected on a channel whose burnout

detection function is enabled, the word “Burnout” is indicated. For all other cases, a numeric

value is displayed.

Over range of analog input channels

• An over range occurs when the measured value of an analog input channel exceeds ±5%

of the measurable range. For example, the measurable range when the measurement range

is 2 V is –2.000 to 2.000 V. If the measured value exceeds 2.200 V, +over range occurs; if the

measured value falls below –2.200 V, –over range occurs.

• For a channel using Linear scaling or Square root, over range occurs if the measured value

falls outside the –5% to 105% range of the specified span. You can also change the setting so

that over range occurs when the measured value falls outside the –5% to 105% range of the

measurable span range. However, +over range occurs if the value excluding the decimal point

exceeds 999999 and –over range if it falls below –999999.

Setup: See page 1-75 in section 1.7.2, “Setting the Over-range Detection Method”.

Bar Graph Display

Measured data is displayed in a bar graph.
Vertical

Channel number, tag, or tag No.

Alarm mark

Alarm point mark

Bar graph (displayed

using the channel display color)

Lower limit of span

Scale

Measured value

Upper limit of span

Unit

Horizontal

Bar graph base position is set to Lower.

Bar graph base position is set to Center.

Bar graph base position is set to Upper.

• Bar Graph Updating

The bar graph is updated at the same interval as numeric values.

1.2 Configuring AI Channels (Analog input (including DI) channels)