5-8

IM 04L51B01-01EN

Module

Electromagnetic relay scanner type modules make measurements by switching mechanical

contact relays on and off.

To ensure that the modules continue to operate reliably and correctly, replace them

periodically as shown below.

Module

Model

Measurement

condition

Replacement

period

Note

Analog input module
(

Electromagnetic relay

scanner type)

GX90XA-10-T1N-3N Continuous use at

measurement interval

1 s

1 year

Continuous use at

measurement interval

2 s

2 years

Continuous use at

measurement interval

5 s

5 years