Performing A/D Calibration and Adjusting the Input Accuracy

1

Tap the

MENU key, Browse tab, Initialize Calibration, and menu A/D calibration.

The A/D calibration call-up screen appears.

2

Tap

Execute.

A screen appears for you to confirm the switch to A/D calibration mode.

3

Tap

OK.

The GX/GP restarts and enters A/D calibration mode.

4

Select the module to calibrate, and tap

Next.

Saves calibration values

Returns the calibration values to their

factory defaults

Returns to the module selection screen

Exits from calibration mode

Calibrate the selected range.

5

For each calibration range, apply the reference values indicated in the table below.

Range

CH1

CH2

20 mV

0 mV (Short circuit)

20 mV

60 mV

0 mV (Short circuit)

60 mV

200 mV

0 mV (Short circuit)

200 mV

1 V

0 V (Short circuit)

1 V

2 V

0 V (Short circuit)

2 V

6 V

0 V (Short circuit)

6 V

20 V

0 V (Short circuit)

20 V

50 V

0 V (Short circuit)

50 V

Pt100-H*

0 Ω (Short circuit)

160 Ω

Pt100*

0 Ω (Short circuit)

400 Ω

* Range calibration of RTDs is not applicable to electromagnetic relay scanner type (Type suffix

code : -T1) analog input modules.

6

Tap

Execute for the range you want to calibrate.

While calibration is in progress, a message indicating this appears. When the calibration is com-

plete, the message “Execution is complete” and the calibration value appear.

7

Repeat step 5 for every range.

8

When all the ranges have been calibrated, tap

Write.

The calibration values will be saved.

To reset the A/D calibration values to their factory defaults, tap

Reset.

You do not have to tap

Write.

9

To calibrate a different module, wire the module to calibrate, tap

Back, and perform

steps 4 to 7.

10

Tap

Exit. A confirmation screen is displayed. Tap OK to exist calibration mode.

Operation complete