Yokogawa Touch Screen GP20 User Manual
Page 439

14
All Rights Reserved. Copyright © 2012-2013, Yokogawa Electric Corporation
<<Contents>> <<Index>>
GS 04L52B01-01EN
May. 7, 2013-00
External Dimensions
M3 screw terminal
107.1(4.22)
26
(1.02)
92(3.62)
100(3.94)
3(0.12)
8(0.31)
(0.07)
1.9
45.1(1.78)
43.8(1.72)
Unit: mm
(approx. inch)
Clamp terminal
107.1(4.22)
26
(1.02)
92(3.62)
100(3.94)
3(0.12)
8(0.31)
(0.07)
1.9
45.1(1.78)
43.8(1.72)
Unit: mm
(approx. inch)
Normal Operating Conditions
Same as the GP main unit
Transport and Storage Conditions
Same as the GP main unit
Effects of Operating Conditions
• Influence of ambient temperature: variation against a
change of 10 °C at an accumulation time of 16.67 ms
or more ± (0.05% of rdg + 0.05%) or below
KpvsAu7Fe, PR20-40: ±(0.05% of rdg + 0.1%) or
below, Cu10Ω: ±(0.2% of rdg + 0.1 °C) or below
No reference contact accuracy is guaranteed.
• Influence of power supply voltage variation:
Accuracy is satisfied in the range of rated power
supply voltage.
• Influence of external magnetic field: Variations
against an AC external magnetic field (50/60 Hz,
400 A/m) are ±(0.1% of rdg+ 0 .1%) or below.