Page 108

102
Tutorial
NIRWare 1.5 Software Manual
5.6.
Using applications and calibrations in Life Cycle state
"Created"
The calibrations and applications can be used in the Life Cycle state 'created'.
This can be helpful to test a calibration during calibration development or to adjust instrument and
application parameters before setting it into the 'checked' state (only Life Cycle 'Electronic Records' or
'Electronic Records'/'Electronic Signatures') or 'approved' state for routine use.
5.6.1. Application for development (Life Cycle state "Created")
For applications in the life cycle state 'created', all application parameters are accessible and can be
modified. This can be used to define application parameters and test them directly with the Operator.
Calibrations can be assigned in the life cycle state 'created' and used for predictions in the Operator
module. Once the application fulfills all requirements for the routine, it will be necessary to secure the
settings and set the calibrations and the application into an 'approved' state.
A test before release can be done with the application in Life Cycle state 'checked' (Life Cycle template
'Electronic Records' or 'Electronic Records'/'Electronic Signatures' only).
NOTE:
Applications or calibrations in the Life Cycle state 'Created' should not be used for routine measurements
at any time, because relevant information that has direct influence on the result and performance of the
method can be modified.
Applications in the Life Cycle state 'Created' have the following attributes:
Only Designers or Administrators have access to applications in the Life Cycle state
'Created'.