BUCHI NIRWare User Manual
Page 170

164
NIRWare Suite
NIRWare 1.5 Software Manual
5. Fiber Optic Liquids
8 scans (Qualitative)
16 scans (Quantitative)
6. Fiber Optic Solids
8 scans (Qualitative)
32 scans (Quantitative)
7. Liquids
8 scans (Qualitative)
16 scans (Quantitative)
8. Solids with Vial Add-On
8 scans (Qualitative)
32 scans (Quantitative)
8. Solids with XL Add-On
8 scans (Qualitative)
32 scans (Quantitative)
8. Solids with Petri Add-On
8 scans (Qualitative)
32 scans (Quantitative)
8. Solids with Tablet Add-On
8 scans (Qualitative)
32 scans (Quantitative)
It is possible to enter a different number of scans optimized to the task.
To set back the default value for the instrument configuration, click on this
button
.
Resolution
Select check box:
Checked = full instrument resolution 8cm
-1
with Boxcar (default)
Unchecked = half resolution (16cm
-1
) with Boxcar (default for Solids
Transmittance)
The default resolution is recommended.
Apodisation 1 for
Phase Correction
(Don't change!)
This parameter is for information only and shows the apodisation function
1 used for phase correction.
A different
may be available in the future (Default
function = Blackman)
Apodisation 2 for
Phase Correction
(Don't change!)
This parameter is for information only and shows the apodisation function
2 used for phase correction.
A different
may be available in the future (Default
function = Blackman)
Apodisation for
Interferogram (Don't
change!)
This parameter is for information only and shows the apodisation function
used for interferogram treatment.
A different
may be available in the future (Default
function = Blackman)
Transfilter
Don't change!
This function is only used during the phase correction.
Checked = Transfilter activated
Unchecked = Transfilter deactivated (default)
Datapoints
Use with care.
Recommendations for lower wavenumber limit 4000cm
-1
and Datapoint
interval 4: