BUCHI NIRWare User Manual
Page 56

50
Tutorial
NIRWare 1.5 Software Manual
Now your calibration is finished and in the approved state, so it can be used for integration into an
application. This is done in the NIRWare Management Console. (Please remember that only applications
with an approved calibration can be approved.)
5.3.6. Integrating a qualitative calibration into an application
The final step in creating an application for identity control with NIR is to integrate the calibration into the
application.
Remember that we started with an application for triplicate measurement of reference substances. This
serves now as a template for the routine application. In effect, we create a copy of the reference method,
make adjustments to this, and integrate the calibration. Adjustments involve mainly the number of
measurements per sample, the SOP text, and Reports. All instrument parameters remain untouched,
ensuring identical spectral conditions for both reference samples and those to be tested.
Both the creation of the routine use application and the integration of the calibration into this is performed
in the NIRWare Management Console (MC).
Load the existing application for data acquisition by choosing ‘Open’ from the ‘Application Designer’ in the
tree list. Then select the application from the list on the right: