BUCHI NIRWare User Manual

Page 84

Advertising
background image

78

Tutorial

NIRWare 1.5 Software Manual

It is not possible to develop only one calibration for determination of both properties. So we need
to create two calibrations, one for each property. First we start with a calibration for Lactose quantification,
so we deselect the unwanted property ‘Fine Sucrose’. This is done by selecting ‘Properties’ in the NIRCal
tree menu, followed by a right-click on the respective property name in the right part of the explorer
window. From the menu, select ‘Remove from Set’ – ‘Calibration Properties’:

Notice the change in the symbol in front of the property name ‘Fine Sucrose’:

Now we are ready for a first calculation

with default settings by clicking on the ‘Overview’ button.

A message appears regarding Calibration- and Validation set selection (C-Set and V-Set):