Sample management, Introduction sample manager – BUCHI NIRWare User Manual
Page 190

184
NIRWare Suite
NIRWare 1.5 Software Manual
NOTE
Make sure that inside an application the Property Names and the SubstanceIDs are unique. The software
will not check this!
NOTE
The start position for the first character is 1.
For not using bar code fractionation set positions to 0.
6.3.
Sample Management
6.3.1. Introduction Sample Manager
The Sample Manager is designed to check and review previous measurement results and system
reports and to enter reference values.
During data acquisition in the Operator, measurement sequences will be recorded. Depending on the
used measurement cell and add-on, a measurement sequence contains 1 sample (NIRMaster,
NIRCheckMaster, Fiber Optic SMA, Fiber Optic Solids, Fiber Optic Liquids, Solids with XL or Petri)
or several samples (Liquids (up to 6), Solids, Vial (up to 6), Tablet (up to 10) and Solid Transmittance).
Each sample contains out of 1 or more spectra depending on the used application setting (Application
Designer - Operator Configuration - Number of Measurement Sequences).
One spectrum is an average of 1 or more scans. A scan records the whole spectral information. The
number of scans used for averaging the spectra depends on the used application setting (Application
Designer - Instrument - Number of Scans).
The following actions are possible in the sample manager:
enter property information (e.g. sucrose or fat [%])
assign properties and reference values to samples
review and print-out of sample reports from routine measurements
review and print-out of system suitability (SST) reports
review and print-out of USP test reports
change sample names
delete unused samples
disable not important routine results (measurement sequences)
electronic signature on batches
verification of signed routine results
electronic signature on results (measurement sequences)