Description of the export fields – BUCHI NIRWare User Manual
Page 282

276
NIRWare Suite
NIRWare 1.5 Software Manual
NOTE:
If the configuration in the LIMS Interface is changed, the Operator and the Control System Service have to
be restarted.
Description of the export fields
The results of the measurements will be validated according the calibration and application settings with
OK if all of the following criteria are fulfilled for Identification (I) and Quantification (Q):
residual is within the given limits (I, Q)
prediction result is within the calibration range (Q)
prediction result is within action and warning limits (Q)
prediction result is within the given distance limit (I)
expected substance matches found substance (I)
Identifier
Data Type (Text:
ASCII or Unicode)
Application Type
(I = Identification,
Q =
Quantification
Remarks
Samplename
Text
I, Q
sample name
DateTime
DateTime
I, Q
UTC Time in the
format: YYYY-mm-
dd HH:MM:SSZ
example:
2006-11-02
14:58:41Z
DateTimeOS
DateTime
I, Q
short date in time
as defined in the
operating system
Date
Date
I, Q
short date as
defined in the
operating system
Time
Time
I, Q
time as defined in
the operating
system
Application
Text
I, Q
application name
Batch
Text
I, Q
batch name
AnalysisID
Text
I, Q
analysis ID
ExpectedSubstance
Text
I
expected
substance identity
SubstanceID
Text
I
ID code for sample
property
FoundSubstance
Text
I
found substance
(equals first entry
in the hitlist; if no
substance was
found, the entry
remains empty)