NIRWare Suite

NIRWare 1.5 Software Manual

NOTE:

If the configuration in the LIMS Interface is changed, the Operator and the Control System Service have to
be restarted.

Description of the export fields

The results of the measurements will be validated according the calibration and application settings with
OK if all of the following criteria are fulfilled for Identification (I) and Quantification (Q):

residual is within the given limits (I, Q)

prediction result is within the calibration range (Q)

prediction result is within action and warning limits (Q)

prediction result is within the given distance limit (I)

expected substance matches found substance (I)

Identifier

Data Type (Text:
ASCII or Unicode)

Application Type
(I = Identification,
Q =
Quantification

Remarks

Samplename

Text

I, Q

sample name

DateTime

DateTime

I, Q

UTC Time in the
format: YYYY-mm-
dd HH:MM:SSZ

example:

2006-11-02
14:58:41Z

DateTimeOS

DateTime

I, Q

short date in time
as defined in the
operating system

Date

Date

I, Q

short date as
defined in the
operating system

Time

Time

I, Q

time as defined in
the operating
system

Application

Text

I, Q

application name

Batch

Text

I, Q

batch name

AnalysisID

Text

I, Q

analysis ID

ExpectedSubstance

Text

I

expected
substance identity

SubstanceID

Text

I

ID code for sample
property

FoundSubstance

Text

I

found substance
(equals first entry
in the hitlist; if no
substance was
found, the entry
remains empty)