Measuring reference spectra – BUCHI NIRWare User Manual
Page 68

62
Tutorial
NIRWare 1.5 Software Manual
To change the settings, first click on the ‘Edit Data Set’ icon
.
Notice the change in the icon bar for Lifecycle
state from ‘Created Idle’ to ‘Created Editing’ (see above).
You can now make changes to all white fields. For our example, we just change the ‘Number of
measurement sequences’ from 1 to 3.
This will cause a triplicate measurement of every reference sample, which creates more variation in the
spectra set later used for calibration. Thus, we will get more robust calibrations.
Save all changes you made to the database by clicking the respective icon
.
The fields are greyed out again afterwards, indicating successful saving.
The method for triplicate reference sample measurement is now still in the ‘Created’ state, where only an
‘Administrator’ can see and use it in the operator software.
To provide this method for the operator user, ‘Put the data set into the next state’ by clicking on the
corresponding icon
.
Now the application is in the ‘Approved’ state.
NOTE
Approved applications cannot be changed anymore. To modify any settings or parameters, create a copy
of the application. The copy will be generated in LC state "created" and can be edited.
The application is now ready for measurement of reference spectra with the Operator software.
5.4.3. Measuring reference spectra
To measure spectra of reference samples, the Operator software is used.