NIRWare Suite

NIRWare 1.5 Software Manual

24 (default)

Successful suitability
test is required for
measurement

A routine measurement can be started only if the SST was successful.

Checked = A routine measurement can be started only

with a successful SST. It is suggested to work with this
adjustment.

Unchecked = A routine measurement can be started

regardless of the SST result (default)

Save Spectra
Interferogram

Next to the spectral information, the interferogram of each measurement
can be stored in the system. Activating this function will increase the use of
database space and the further use of the interferogram data is not yet
implemented.

Checked = save spectra interferogram is activated

Unchecked = save spectra interferogram is deactivated

(default)

Gain of ADC Detector
for samples

This parameter is for information only.
The detector has 5 levels of sensitivity. Depending on the measurement cell
used, the optimal gain is used automatically.

Gain of ADC Detector
for references

This parameter is for information only.
The detector has 5 levels of sensitivity. Depending on the measurement cell
used, the optimal gain is used automatically.

External Reference
Mode

Mode to measure the external reference. Choose from the following
possibilities:

Interval or Temperature Drift = regularly or if the temp. changes

Manual = Manual reference measurement

Measurement Sequence = at each start of a measurement sequence

Once a day = once a day, before the first measurement

Interval = regularly (see interval value below)

Temperature Drift = after a certain temperature drift of the

interferometer

(see value below)

Batch = once after a new batch has been created

Disabled = No measurement for computing

Default:

1. NIRMaster

Interval (60 min)

2. Fiber Optic SMA (for
continuous measurement)

Manual (60 min)

3. CheckMaster Transmittance

Measure Sequence

4. Solids Transmittance

Measure Sequence

5. Fiber Optic Liquids

Interval (60 min)

6. Fiber Optic Solids

Interval (60 min)

7. Liquids

Measure Sequence

8. Solids with Vial Add-On

Measure Sequence