Resetting the lifetime counter of lamp or laser, Resetting lifetime counter of lamp or laser – BUCHI NIRWare User Manual
Page 331

NIRWare 1.5 Software Manual
NIRWare Operator
325
As soon as the primary lamp has been replaced, the described steps should be followed and the system
should be switched back to the primary lamp. For this purpose, open the following dialog and click the
button Activate Primary Lamp.
To reset the Primary lamp counter, see also
Resetting the lamp lifetime counter
7.8.4. Resetting the lifetime counter of lamp or laser
Menu:
NIRWare Operator / Advanced / Service / Assembly Exchange
Lamp (Laser)
Icon:
After the installation of a new lamp or laser the corresponding lifetime counter should be reset.
To reset the lifetime counter of the primary lamp or the laser, open the Service module and select
"Assembly Exchange Lamp" respectively Assembly Exchange Laser.