NIRWare 1.5 Software Manual

NIRWare Suite

197

The property is now added to the samples and the classification information can be entered directly by
clicking on the check box or use "Enter" and the arrow down button to fill in the information.

After all entries are made, click on

to save changes to the database.

With two exceptions, properties can also be removed from samples.

To remove a property, select the corresponding entry in "Other Assigned Properties" and click

on

. The entered reference values will be lost.

In case the sample is used in a calibration the property and the reference values can not be deleted. The
following message appears:

Only properties added to the sample manually can be removed. Properties related to the routine
application can not be removed.

The colour of the row indicator shows the type of property:

Property from routine application. This property will be assigned automatically to samples

measured with a routine application. These properties can not be removed from samples.

Property assigned by user. This property can be removed from samples as long as the

corrwesponding samples are not used in a calibration.