See also "Definition of calibration range, warning and action limits".

Default: lower limit of calibration range

To set back the default value for the action limit, click on this button

.

Bias

Linear correction of a predicted result using bias function. Correction will be
displayed in report.

Default: 0

Slope

Linear correction of a predicted result using slope function. Correction will
be displayed in report

Default: 1

Label Claim (CUT)

Content of active pharmaceutical ingredient per sample

L1 (CUT)

Maximum allowed acceptance value

L2 (CUT)

Maximum allowed range for deviation of each dosage unit tested from the
calculated value of M.

T (CUT)

Target test sample amount at time of manufacture.

* non bold parameters are advanced parameters.

6.2.7. Definition of calibration range, warning and action limits

The calibration range of a quantitative calibration is defined by the data (minimum and maximum values
in calibration set) used for calibration development. By default the related warning and action limits are set
to their maximum range which is identical to the calibration range.
It will be possible to define warning and action limits within the calibration range by entering the
corresponding values in the parameter list.