BUCHI NIRWare User Manual
Page 163

NIRWare 1.5 Software Manual
NIRWare Suite
157
See also "Definition of calibration range, warning and action limits".
Default: lower limit of calibration range
To set back the default value for the action limit, click on this button
.
Bias
Linear correction of a predicted result using bias function. Correction will be
displayed in report.
Default: 0
Slope
Linear correction of a predicted result using slope function. Correction will
be displayed in report
Default: 1
Label Claim (CUT)
Content of active pharmaceutical ingredient per sample
L1 (CUT)
Maximum allowed acceptance value
L2 (CUT)
Maximum allowed range for deviation of each dosage unit tested from the
calculated value of M.
T (CUT)
Target test sample amount at time of manufacture.
* non bold parameters are advanced parameters.
6.2.7. Definition of calibration range, warning and action limits
The calibration range of a quantitative calibration is defined by the data (minimum and maximum values
in calibration set) used for calibration development. By default the related warning and action limits are set
to their maximum range which is identical to the calibration range.
It will be possible to define warning and action limits within the calibration range by entering the
corresponding values in the parameter list.