Measuring reference spectra – BUCHI NIRWare User Manual
Page 33

NIRWare 1.5 Software Manual
Tutorial
27
You can now make changes to all white fields. For our example, we
just change the ‘Number of
measurement sequences’ from 1 to 3.
This will cause a triplicate measurement of every reference sample, which creates more variation in the
spectra set later used for calibration. Thus, we will get more robust calibrations.
Save all changes you made to the database by clicking the respective icon
.
The fields are greyed out again afterwards, indicating successful saving.
The method for triplicate calibration sample measurement is now still in the ‘Created’ state, where only a
‘Designer’ or ‘Administrator’ can see and use it in the operator software.
To provide this method for the operator user, ‘Put the data set into the next state’ by clicking on the
corresponding icon
.
Now the application is in the ‘Approved’ state.
NOTE
Approved applications cannot be changed anymore. To adjust something, you first have to create a copy
of the application, which again starts in the ‘created’ state. Then you can apply changes to this new
application.
The application is now ready for measurement of reference spectra with the Operator software.
5.3.3. Measuring reference spectra
To measure spectra of reference samples, the Operator software is used.
Measurement of reference spectra is the second step in the development process.