To save the calibration to the database, click on the ‘save’ icon in the menu bar

.

Type in a name for your NIRCal project and click on the ‘Save As’ icon.

So far the

calibration is still in the ‘created’ state. If you would try to use it in an application, this would

prevent you from changing your application (!) into the next state, because an approved application
requires an approved calibration.

It is suggested to recalculate the first calibration. This is a master calibration, which organises the spectra
in the project and can never be put the approved LC state. Copy this calibration, open the copy (LC: Edit),
rename this: give a short name preferable with date at the end. Make a new calculation after renaming.
Put this calibration to "approved" state.

To put your application into the next state, right-click on the name of the calibration in the NC Explorer,
select ‘Lifecycle’ and ‘Next’.