For both report formats, for a material to be successfully identified, a green “OK” should
appear next to the Name, Distance, and Spectral Residual in the area of the report
corresponding to the calibration that identifies the material just scanned. If one of the 3
criteria is not fulfilled, the substance is not identified. Only if Distance and Residual are
OK, there is a substance name in the "Found substance" field. Reports may be printed by
clicking the print icon above the report.

In order to view the spectrum of the material just scanned, the user clicks the “spectra”
tab on the right side of the screen. If reference scans are being displayed, it will be
necessary to “uncheck” these scans under the plots to allow the y-axis to scale
appropriately.