Library parameters – BUCHI NIRWare User Manual
Page 287

NIRWare 1.5 Software Manual
NIRWare Suite
281
The library will now be added to the tree view on the left side under the menu Open and three additional
submenus are assigned to it: Spectra, Validate Library, Library Test.
NOTE
The library can only be saved when a library name and a template spectrum was defined.
Library parameters:
Library name
Name of the library (mandatory field)
Library description
Optional description of the library
Library type
Type of library, preset to static
Method formula
Select method from the drop-down list:
Maximum Distance of 2nd Derivative
Maximum Distance
Wavelength Correlation of 2nd Derivative
Wavelength Correlation
Squared Differences of 2nd Deriv with Smoothing
Squared Differences of Normalisation 0 to 1
Squared Differences of Vector Normalisation
Correlation Coefficient
Euclidian Distance
Squared Difference of Derivative
Difference of Derivative
Squared Differences
Difference
Wavelength selection
Wavelength range in which the spectra are compared
Distance limit
Limit value defining the maximum allowed distance between a
given/measured spectrum and the most similar spectrum within the
library. The value depends on the method formula selected for
prediction and is defined in the library parameters.
Validation type
Select between internal or external validation from the drop down menu
Minimum measurement cell
temperature
Minimum temperature the measuring cell may have (valid only for liquid
cells)
Measurement temperature
maximum
Maximum temperature the measuring cell may have (valid only for
liquid cells)