Carrying out a measurement with the operator – BUCHI NIRWare User Manual
Page 116

110
Tutorial
NIRWare 1.5 Software Manual
where the first internal reference was recorded right before the external reference. Internal referencing
minimizes the need for (inconvenient) user interaction in situations where no automatic placement of the
reference is possible.
NOTE
Different modes for reference measurements are available. Batch, Manual, Temperature Drift, Once a
day, Measure Sequence, Interval and Disabled.
In case you plan to switch between different applications and want to avoid external reference spectra
during each change, it
might be useful to set the external referencing to “Manual” or “Once a day”.
It is not recommended to disable the external referencing as that means that the spectrum is not
taken into account for the calculation according to the above formula.
5.11. Carrying out a measurement with the Operator
The measurement sequence should start with the N-500 warmed up for at least 30
minutes and with reference cap in place on the probe. The NIRWare Operator software is
started by double-clicking the respective icon, or from the Windows Start-Menu. The window
below appears:
The user then types in his user name and password in the fields indicated