BUCHI NIRWare User Manual

Page 110

Advertising
background image

104

Tutorial

NIRWare 1.5 Software Manual

NOTE:

If you modify a calibration in life cycle state 'created' that was used for predictions already, all previous
measurement sequences and result reports generated with that calibration will be modified according to
the new calibration setting.

To update the calibration in the application, you need to de-select the calibration and re-asign the
calibration new to the corresponding property or property group.

After an application has been tested and is ready to be released for routine use, the following steps have
to be followed:

the used calibrations have to be set to life cycle state 'approved'

the application has to be set to life cycle state 'approved'

all generated measurement sequences during testing and development should be

archived.

5.6.2. Application in routine use (Life Cycle state 'Approved')

For routine measurements it is necessary to ensure the traceability between results, spectra,
measurement methods and calibrations. Therefore applications and calibrations for routine use can not be
modified anymore.

An application or calibration can be locked by setting them to the Life Cycle state 'Approved'. With the Life
Cycle template 'Electronic Records' or 'Electronic Records'/'Electronic Signatures', a test of the
application before release in approved state can be done by using the intermediate Life Cycle state
'checked'.

To bring an application in a higher Life Cycle state, the linked calibrations have to be put into approved
Life Cycle state first.

An application for routine use in Life Cycle state 'approved' has the following attributes:

The access for the user group 'Operators' is limited to applications in 'approved' state for
routine use.