NIRWare 1.5 Software Manual

Tutorial

107

5.8.1. Manual SST measurement

1. Start NIRWare Operator

2. select an Application from the DB

3. In the menu under Advanced select "Perform Suitability Test."

Or start NIRWare Operator, use "Open service tool" and choose "Suitability test".

After completion of the System Suitability Test a Report is displayed on the right side of the Operator
window. A detailed Report can be viewed in the NIRWare Sample Manager.

5.8.2. Automatic SST measurement

By default, a SST will be measured automatically before the first sample measurement in an application, if
the set time limit (every 24h) is exceeded.

The time interval can be set in the

Application Designer

.

5.9.

Measuring a NADIA

NADIA stands for „NIRWare Automatic Diagnose“ and is a service tool to examine the basic functions of
the BUCHI NIRFlex-N500.

It will be used from Service Technicians to evaluate malfunctions of the instrument or to do a performance
check of the instrument as part of a service contract.

The NADIA is started via the Windows Start Menu (Start\All programs\ BUCHI\ NIRSolutions\ Tools\
Nadia)