The tables comprise the following columns:
Sample Name: Date and time the sample was measured and name of the sample
Property Name: Name of the property
Time Stamp: Date and time the sample was measured
Scans: Number of scans with which the sample was measured
Resolution: Resolution with which the sample was measured
Comment: Optional comment to the sample
Spectrum ID: ID of the sample spectrum
To be able to assign spectra to your library, you first have to set the corresponding library to the Lifecycle
state "Editing". For this purpose, click the Edit DataSet icon
. The blue arrow buttons in the center of
the window will now become active.
You can now highlight the row of the sample(s) you want to add to the library and click the left arrow
button. The sample is moved from the lower to the upper table.
To remove a spectrum from the library again, highlight the corresponding row and click the right arrow
button.
To check whether a spectrum assigned to the library meets the requirements defined in the library
parameters, click Check. The check results, i.e. the Minimum external distance, the Maximum internal
distance and the Maximum allowed distance, will be displayed in the lower part of the window.
NOTE:
Make sure, that only spectra are used to built up a library that have the same wavenumber range and
number of data points (e.g. 10000cm
-1
- 4000cm
-1
; 1501 data points).
To create a new library from spectra contained in the currently active library, highlight the corresponding
spectra in the upper table and click New Library. The following window opens: