Using bar codes – BUCHI NIRWare User Manual

Page 323

Advertising
background image

NIRWare 1.5 Software Manual

NIRWare Operator

317

If the test passes, the Operator loads the job, starts the corresponding application and fills in the selected
fields in the input mask. All fields activated for LIMS should be also activated in "Operator Configuration
- Sample Description Fields"
.

To start the measurement, click the "Start" button

.

During the measurement the import file is locked.

If the measurement is interrupted no export file will be generated.

If the measurement is finished successfully, but also in case the measurement is out of specification or
the expected substance does not match, and after signing the measurement (depending on the lifecycle
state), the result files will be written to the outbox. There is a separate file for each sample. The import file
will be deleted automatically.

NOTE:

An export file is only generated if the import file could be read properly.

In the Application Designer it is recommended to select the options "Field Reset" and "Read Only", so that
the input fields are cleared after the measurement.

If the configuration in the LIMS Interface is changed, the Operator and the Control System Service have to
be restarted.

7.7.

Using bar codes

After the configuration of application and property-IDs, we are ready to use the NIRWare Operator with
the bar code reader:

After loading the application, we read

the bar code into the ‘Batch’-field: