NIRWare 1.5 Software Manual

NIRWare Suite

167

8. Solids with XL Add-On

Interval (20 min)

8. Solids with Petri Add-On

Interval (60 min)

8. Solids with Tablet Add-On

Measure Sequence

External Reference
Interval

Interval in minutes to measure the external reference, if the mode to
measure the

external reference was selected as “interval”.

Internal Reference
Mode

Mode to measure the internal reference (not available for Solids
Transmittance). Choose from the following possibilities:

Disabled = No measurement for computing

Manual = Manual reference measurement

Measurement Sequence = at each start of a measurement sequence

Interval = regularly (see interval value below)

Temperature Drift = after a certain temperature drift of the

interferometer (see value below)

Batch = once after a new batch has been created

Default:

1. NIRMaster

Interval (10 min)

2. Fiber Optic SMA (for
continuous measurement)

Interval (10 min)

3. CheckMaster Transmittance

n.a.

4. Solids Transmittance

n.a.

5. Fiber Optic Liquids

Interval (10 min)

6. Fiber Optic Solids

Interval (10 min)

7. Liquids

Disabled

8. Solids with Vial Add-On

Disabled

8. Solids with XL Add-On

Disabled

8. Solids with Petri Add-On

Interval (10 min)

8. Solids with Tablet Add-On

Disabled

Internal Reference
Interval

Interval in minutes to measure the internal reference, if the mode to
measure the internal reference was selected as “interval”.

Max. Drift for
Reference
Temperature

Maximum allowed drift in °C inside the NIRFlex N-500. If this maximum drift
is exceeded, the software will ask to do an SST and to measure external
and internal reference.

Sample holder rotation
for Petri dish

Special parameter for Measurement Cell

“NIRMaster” and “Solids with Petri

add-on

” to turn on or off the rotation.

Checked = rotation of petri dish add-on is turned on

Unchecked = rotation of petri dish add-on is turned off

Heater Temperature
Nominal

Special parameter for Measurement Cell Liquids.
Set temperature in [°C] in cuvette holder for liquid measurements.

Default: 35°C
Range: ambient temperature (+10°C) to 65°C

Sample Gain

Setting for NIRFlex Solids Transmittance.