Modular option c – West Control Solutions KS98-1 User Manual

Page 15

Advertising
background image

Modular Option C

Each module has two channels which can be configured independently.

A/D-Converter

Resolution: 20,000 (50Hz) or 16,667 (60Hz) steps for the selected
measuring range.

Conversion time: 20ms (50Hz) or 16.7ms (60Hz).

D/A-Converter

Resolution: 12 Bit

Refresh-Rate: 100 ms

Cut-off frequency

Analog: fg=10Hz

Digital: fg=2Hz

Measurement cycle: 100 ms per module

R_INP Resistive input module

(9407-998-0x201)

Connection method: 2, 3 or 4-wire circuit (with 3 und 4-wire con-
nection, only one channel can be used).

Sensor current:

£

0,25mA

Resistance thermometer

Type

Range °C

Overall error

Resolution K/Digit

Pt100

-200...850°C

£ 2 K

0,071

Pt100

-200...100°C

£ 2 K

0,022

Pt1000

-200...850°C

£ 2 K

0,071

Pt1000

-200...100°C

£ 2 K

0,022

Ni100

-60...180°C

£ 2 K

0,039

Ni1000

-60...180°C

£ 2 K

0,039

Linearization: in °C oder °F

Lead resistance
Pt (-200...850°C):

£ 30

W

per lead

Pt (-200...100°C), Ni:

£ 10

W

per lead

Lead resistance compensation
3 and 4-wire connection: not necessary.

2-wire connection: compensation via the front with short-circuited
sensor. The calibration values are stored in a non-volatile memory.

Lead resistance effect
3 and 4-wire connection: negligible

Sensor monitoring
Break: sensor or lead. Short-circuit: triggers at 20 K below mea-
suring range.

Resistance / Potentiometer

Range R

ges

/

W

Overall error

Resolution

W/Digit

0...160

W

£ 1%

0,012

0...450

W

£ 1%

0,025

0...1600

W

£ 1%

0,089

0...4500

W

£ 1%

0,025

Characteristic: resistance-linear

Lead resistance or 0%/100% compensation:
via the front with short-circuited sensor. The calibration values are
stored in a non-volatile memory.

w

Variable resistance (only 2-wire connection): compensation for 0%

w

Potentiometer: compensation for 0% and 100%

Lead resistance effect
3 and 4-wire connection: negligible

Sensor monitoring:
break: sensor or lead

TC_INP Thermocouple, mV, mA Module

(9407-998-0x211)

Thermocouples

To DIN IEC 60584 (except L, W/C, D)

Type

Range

Overall

error

K/Digit

L

-200...900°C

£ 2 K

0,080

J

-200...900°C

£ 2 K

0,082

K

-200...1350°C

£ 2 K

0,114

N

-200...1300°C

£ 2 K

0,129

S

-50...1760°C

£ 3 K

0,132

R

-50...1760°C

£ 3 K

0,117

B

(1)

(25) 400...1820°C

£ 3 K

0,184

T

-200...400°C

£ 2 K

0,031

W(C)

0...2300°C

£ 2 K

0,277

D

0...2300°C

£ 2 K

0,260

E

-200...900°C

£ 2 K

0,063

*

(1) Values apply from 400°C

Linearization: in °C or °F
Linearity error: negligible

Input resistance:

³

1M

W

Temp. compensation (CJC): built in

Error:

£

0.5K/10K

External CJC possible: 0...60 °C or. 32...140 °F

Source resistance effect: 1mV/k

W

Sensor monitoring:
Sensor current:

£

1

m

A

Reversed polarity monitor: triggers at 10K below measuring range

mV input

Range

Overall error

Resolution

0...30 mV

£ 45 mV

1,7 mV

0...100 mV

£ 150 mV

5,6 mV

0...300 mV

£ 450 mV

17 mV

Input resistance:

³

1M

W

Break monitoring: built in. Sensor current:

£

1

mA

9499-040-82711

Technical data

I-15

Advertising